EIfClip™ Absorbable Laparoscopic Hemoclips from China Suppliers | High-Quality Medical Factory Products
PRODUCT FEATURES
The raw material of the inner layer is preferably poly-dioxanone, which has good hydrophilicity and softness. No acute angle to cause blood vessels damage or other tissue hazards, and the degradation rate is consistent with the outer layer.
The inner and outer layers are designed in two distinct colors. The effect of clipping the tube cavity of the inner and outer layer is clear at a glance, avoiding misjudgment in actual using.
In actual surgery, the number of lumens to be ligated often appears to be singular. Single packaging is more economical and reasonable to reduce unnecessary medical expenses for patients.
PRODUCT DESCRIPTION
| Product Use: | For ligation of cystic duct, arterial vessel with diameter of 5 mm or below. | Product Highlights: | With a dual-layer structure, complete degradation and absorption in 39 weeks, degradation products of water and carbon dioxide. |
| Device Classification: | [CN] Class III | Material: | PGA, PDO |
| Specification Description: | Different specifications represent 1, 2 and 3 clips in a package, respectively | Storage Conditions: | The product shall be stored in an environment that is cool, dry, well-ventilated, clean and free from corrosive gases with the temperature of no more than 40℃. |
Product Specification
| Product name | EIfClipTM Absorbable Hemostatic Ligation Clips |
| Instructions | Open the paper-plastic package, take out the product and use it with the special clip applier. |
| Country of origin | China |
| Structure | Crocodile mouth shape with double parts |
| Package |
1pc/bag,2pcs/bag,3pcs/bag(sterilized package) Single sterile package, used on demand to avoid secondary contamination; Used scientifically and economically to reduce unnecessary medical expenses. |
| Retention in the body | Degradable and absorbable |
| Product structure and composition | Consists of four parts, namely inner card, outer card, barrel and plunger, in which the inner card and outer card are implantable parts, and the barrel and plunger assist the inner card and outer card to be implanted into human body. The material of outer part is PGA, and the material of inner part is PDO. |
| Visual identity | The inner and outer layers are designed in two distinct colors. The effect of clipping the tube cavity of the inner and outer layers is clear at a glance, avoiding misjudgment in actual using. |
| Product design | The raw material of the inner part is poly-dioxanone, which has good hydrophilicity and softness. It has no acute angle, which could avoid blood vessels damage or other tissue hazards. The two arms of the inner clip are provided with alternating rack, which is conducive to increasing the friction between the inner clip and the tubular tissue, so that the clamping is more firm and less falling off occurs. |
| Precautions |
The product is for single use; it is not designed to be used in case of damaged package. The product must not be used by unskilled or untrained personnel. |
| Shelf life | 2 years |
Schematic Diagram of Tubular Tissue Ligation
Absorption rate curves of outer and inner circlips products:
Frequently Asked Questions (FAQ)
Q: What materials are the EIfClip™ Absorbable Hemostatic Ligation Clips made of?
The ligation clips are designed with a dual-layer structure where the outer part is made of Polyglycolic Acid (PGA) and the inner part is made of Polydioxanone (PDO).
Q: How long does it take for the clips to degrade and absorb in the body?
The EIfClip™ ligation clips are completely degraded and absorbed in the body within 39 weeks. The degradation products are naturally metabolized into water and carbon dioxide.
Q: What are the primary surgical uses and size limitations for these clips?
They are used for the ligation of the cystic duct and arterial vessels with a diameter of 5 mm or below.
Q: Why are the inner and outer layers designed in two different colors?
The distinct two-color design allows for clear visual identification of the inner and outer parts. This makes the clipping effect on the tube cavity immediately clear, preventing misjudgment during surgical procedures.
Q: How should the EIfClip™ ligation clips be stored?
The product must be stored in a cool, dry, well-ventilated, and clean environment free from corrosive gases, at a temperature not exceeding 40℃.
Q: Can these ligation clips be reused or resterilized?
No, the product is strictly for single use. It must not be reused, and it should not be used if the sterile packaging is damaged.